Usefulness of volumetric BMD measurement by using low dose CT image acquired on L-spine Bone SPECT/CT

Department of Nuclear Medicine, Asan Medical Center, Seoul, Korea

Correspondence to Hyun Soo Ko, Department of Nuclear Medicine, Asan medical Center, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, Republic of Korea, Tel :+82-2-3010-4601, E-mail : hyunsoo87@amc.seoul.kr

Volume 27, Number 2, Article 16, November 2023. Korean J Nucl Med Technol 2023;27(2):16. https://doi.org/10.12972/kjnmt.20230016
Received on August 01, 2022, Revised on August 23, 2023, Accepted on September 12, 2023, Published on November 31, 2023.
Copyright © 2023 Author(s). This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: CT scan makes up for the weak point of the nuclear medicine image having a low resolution and also were used for attenuation correction on image reconstruction. Recently, many studies try to make use of CT images additionally, one of them is to measure the bone mineral density(BMD) using Quantitative CT(QCT) software. BMD exams are performed to scan lumbar and femur with DXA(Dual-Energy X-Ray Absorptiometry) in order to diagnose bone disease such as osteopenia, osteoporosis. The purpose of this study is to identify the usefulness of QCT_BMD analyzed with low dose CT images on L-spine Bone SPECT/CT comparing with DXA_BMD. Materials and Methods: Fifty five women over 50 years old (mean 66.4 ± 9.1) who took the both examinations(L-spine Bone SPECT/CT with SIEMENS Intevo 16 and DXA scan with GE Lunar prodigy advance) within 90 days from April 2017 to July 2022, BMD, T-score and disease classification were analyzed. Three-dimensional BMD was analyzed with low dose CT images acquired on L-spine Bone SPECT/CT scan on Mindways QCT PROTM software and two-dimensional BMD was analyzed on DXA scan. Basically, Lumbar 1-4 were analyzed and the patients who has lesion or spine implants on L-spine were excluded for this study. Pearson’s correlation analysis was performed in BMD and T-score, chi-square test was performed in disease classification between QCT and DXA. Results: On 55 patients, the minimum of QCT_BMD was 18.10, maximum was 166.50, average was 82.71 ± 31.5 mg/cm3. And the minimum of DXA-BMD was 0.540, maximum was 1.302, average was 0.902 ± 0.201 g/cm2, respectively. The result shows a strong statistical correlation between QCT_BMD and DXA_BMD(p<0.001, r=0.76). The minimum of QCT_T-score was -5.7, maximum was -0.1, average was -3.2 ± 1.3 and the minimum of DXA_T-score was -5.0, maximum was 1.7, average was -2.0 ± 1.3, respectively. The result shows a statistical correlation between QCT T-score and DXA T-score (p<0.001, r=0.66). On the disease classification, normal was 5, osteopenia was 25, osteoporosis was 25 in QCT and normal was 10, osteopenia was 25, osteoporosis was 20 in DXA. There was under-estimation of bone decrease relatively on DXA than QCT, but there was no significant differences statistically by chi-square test between QCT and DXA. Conclusion: Through this study, we could identify that the QCT measurement with low dose CT images QCT from L-Spine Bone SPECT/CT was reliable because of a strong statistical correlation between QCT_BMD and DXA_BMD. Bone SPECT/CT scan can provide three-dimensional information also BMD measurement with CT images. In the future, rather than various exams such as CT, BMD, Bone scan are performed, it will be possible to provide multipurpose information via only SPECT/CT scan. In addition, it will be very helpful clinically in the sense that we can provide a diagnosis of potential osteoporosis, especially in middleaged patients.
Keywords

L-spine SPECT/CT, BMD, QCT, DXA