How Auto-Segmentation Errors Turn Osteopenia into PseudoOsteoporosis: A Case of Tissue Area Loss and Neutral Zone OverSubtraction

Department of Nuclear Medicine, Jeonbuk National University Hospital, Jeonju, Korea

Correspondence to Han-Kyung Seo. Department of Nuclear Medicine, Jeonbuk National University Hospital. 20, Geonji-ro, Deokjin-gu, Jeonju, 03080, Republic of Korea. Tel: +82-63-250-2321, E-mail: shg@jbuh.co.kr

Volume 30, Number 1, Article 7, May 2026. Korean J Nucl Med Technol 2026;30(1):7. https://doi.org/10.12972/kjnmt.2026.30.1.7
Received on March 17, 2026, Revised on April 21, 2026, Accepted on April 21, 2026, Published on May 31, 2026.
Copyright © 2026 Author(s). This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

In dual-energy X-ray absorptiometry, accurate identification of the soft tissue area is essential for establishing baseline attenuation and ensuring precise bone mineral density measurement. This study analyzes a 75-year-old female patient who demonstrated an abrupt decline in femoral T-scores within one year. Daily quality assurance of the device and operator precision (least significant change: 0.017 g/cm²) were within normal limits. However, due to an auto-segmentation error, the soft tissue area was lost (area = 0), and a high-density neutral zone was misclassified as the background reference, resulting in excessive subtraction of bone signal. Although the original dataset was immediately deleted due to clinical workflow demands, limiting detailed retrospective analysis, a reproduction experiment using a volunteer successfully demonstrated the same algorithmic error mechanism.
Keywords

Dual-energy X-ray Absorptiometry, Neutral Zone, Soft Tissue Area, Over-subtraction, Auto-segmentation Error, Pseudo-osteoporosis