Korean Journal of Nuclear Medicine Technology (Korean J Nucl Med Technol)

Open access, Peer Reviewed

Indexed in KCI, DOAJ

pISSN 1229-9901
eISSN 2982-8406

Original article

Evaluation of Spatial Resolution in CZT-Based vs NaI(Tl)-Based SPECT/CT

1Department of Nuclear Medicine, Seoul National University Hospital
2Department of Nuclear Medicine, Seoul National University College of Medicine
3Interdisciplinary Program in Bioengineering, Seoul National University

Correspondence to Jung Jin Lim, Department of Nuclear Medicine, Seoul National University Hospital, 101, Daehak-ro, Jongno-gu, Seoul, 03080, Republic of Korea. Tel: (02) 2072-2508, Fax: (02) 747-0208, E-mail: nmpotato@nate.com

Volume 30, Number 2, Article 25, November 2026. Korean J Nucl Med Technol 2026;30(2):25. https://doi.org/10.12972/kjnmt.2026.30.2.25
Received on October 26, 2026, Revised on November 16, 2026, Accepted on November 17, 2026, Published on November 30, 2026.
This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: SPECT/CT systems are shifting from NaI(Tl) scintillation detectors to CZT semiconductor detectors, yet quantitative comparisons of their ability to resolve closely adjacent organs or lesions remain limited. This study compared the spatial resolution and source-separation performance of CZT-based and NaI(Tl)-based SPECT/CT by line-profile analysis at varying distances between two adjacent sources. Materials and Methods: A CZT-based VERITON-CT (Spectrum Dynamics, Israel) and a NaI(Tl)-based Discovery NM/CT 670 (GE Healthcare, USA) were used. A 50 ㎖ syringe was filled with 99mTcO₄⁻ at 9,250 k㏃/㎖ and a 5 ㎖ syringe at 1,850 k㏃/㎖ (concentration ratio 1:0.2). Images were acquired with gaps of 0, 3, 5, 10, and 20 ㎜ between the syringes and reconstructed by OSEM with CT-based attenuation correction according to each system’s clinical protocol. Line profiles through the centers of both syringes were used to measure the full width at half maximum (FWHM), contrast ratio (CR), and valley-to-peak ratio (VPR). Results: Neither system separated the sources at 0 or 3 ㎜. The VPR referenced to the 5 ㎖ peak was 94.91% (CZT) vs 43.66% (NaI(Tl)) at 5 ㎜ and 43.63% vs 21.51% at 10 ㎜, but 2.99% vs 10.43% at 20 ㎜. The FWHM of the 5 ㎖ syringe was consistently narrower with CZT: 7.98 vs 11.60 ㎜ at 10 ㎜ and 10.68 vs 13.35 ㎜ at 20 ㎜. The FWHM of the 50 ㎖ syringe was stable across distances (CZT 25.66–25.90 ㎜; NaI(Tl) 24.00–24.43 ㎜). Conclusion: CZT-based SPECT/CT yielded a narrower FWHM for the small source with a short acquisition and cleaner separation at 20 ㎜, whereas NaI(Tl)-based SPECT/CT separated the sources better at 5–10 ㎜. Because this difference may reflect reconstruction and post-filtering settings rather than the detectors alone, system selection and reconstruction optimization should follow the purpose of each examination rather than assuming one system is superior.
Keywords

CZT Detector, NaI(Tl) Detector, SPECT/CT, Spatial Resolution, FWHM, Valley-to-Peak Ratio

Section