Comparative Analysis of Dynamic and Static Salivary Gland Scintigraphy Using Ejection Fraction

Deparment of Nuclear Medicine, Kyunghee University Hospital at Gangdong, Seoul, Korea

Correspondence to Gi-Woong Jun, Department of Nuclear Medicine, Kyunghee University Hospital at Gangdong, 892 Dongnam-ro, Gangdong-gu, Seoul, 05278, Republic of Korea. Tel :+82-2-440-6892, E-mail : wrd9477@khnmc.or.kr

Volume 29, Number 1, Article 7, May 2025. Korean J Nucl Med Technol 2025;29(1):7. https://doi.org/10.12972/kjnmt.2025.29.1.7
Received on February 24, 2025, Revised on May 18, 2025, Accepted on May 19, 2025, Published on May 31, 2025.
Copyright © 2025 Author(s). This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: This study aims to compare the ejection fraction (EF%) values derived from static and dynamic salivary gland scintigraphy and assess whether there are statistically significant differences between the two imaging techniques. Materials and Methods: Thirty patients (10 males, 20 females; mean age: 65.9 years) complaining of xerostomia underwent dynamic salivary scintigraphy with 185 MBq (5 mCi) of 99mTc-pertechnetate. Imaging was performed using a Discovery NM630 gamma camera (GE Healthcare, USA). Vitamin C powder (2 g) was administered as a stimulant at 40 minutes after radiotracer injection. Dynamic images were acquired at 60-second intervals for 50 minutes. Static images were reconstructed by summing specific frame sets pre- and poststimulation. Ejection fraction was calculated using the Xeleris Functional Imaging Workstation (GE Healthcare, USA), with regions of interest applied to the right parotid and submandibular glands. Statistical analysis was conducted using paired t-tests and correlation coefficients. Results: The ejection fraction average was 81.94±7.94% when calculated using a dynamic method and 80.94±7.95% when calculated using a static method, which was larger when calculated using a dynamic method by a difference of 0.99±3.21% and the correlation was 0.918, and there was no statistically significant difference(P > 0.05). Conclusion: The scan time and equipment operation time are different according to the dynamic and static methods, and each method has advantages and disadvantages. As designating the stimulation time or image composition time, there was a limitation in deriving results according to retrospective analysis, but in this study, since there is no significant difference in the ejection fraction according to method, it is considered that the scan will proceed more efficiently if a more efficient method is selected and used for each laboratory according to the situation among the two methods.
Keywords

Salivary gland scintigraphy, Static method, Dynamic method, Ejection fraction