Original article

Differences and Agreement by Background Measurement Site in 99mTc-Pertechnetate Thyroid Uptake

Department of Nuclear Medicine, Gachon University Gil Medical Center, Incheon, Korea

Correspondence to Wang-Hui Lee, Department of Nuclear Medicine, Gachon University Gil Medical Center. 21, Namdong-daero 774beon-gil, Namdong-gu, Incheon, 21565, Republic of Korea. Tel: +82-32-460-3310, E-mail: lwh840614@gilhospital.com

Volume 30, Number 2, Article 11, November 2026. Korean J Nucl Med Technol 2026;30(2):11. https://doi.org/10.12972/kjnmt.2026.30.2.11
Received on May 11, 2026, Revised on May 31, 2026, Accepted on June 01, 2026, Published on November 30, 2026.
This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: This study aimed to compare thyroid uptake values according to background measurement site (mid-thigh versus neck using a B-type filter) in 99mTc-pertechnetate thyroid uptake studies, evaluate the agreement between the two methods, and assess the influence of sex-related differences. Materials and Methods: A retrospective analysis was performed in 98 patients who underwent 99mTc-pertechnetate thyroid uptake studies. Background activity was measured at both the neck and mid-thigh under identical geometric conditions. Thyroid uptake values were calculated using each background method. Paired t-tests, Pearson correlation analysis, and Bland–Altman analysis were performed to compare the two methods. Sex-based differences in thyroid uptake values (thigh − neck) were additionally evaluated using an independent samples t-test. Results: Neck background activity measured using a B-type filter was significantly higher than mid-thigh background activity (21,472.2±4,268.8 cpm vs. 18,660.3±3,587.4 cpm, P < 0.001). Thyroid uptake calculated using thigh background was also significantly higher than uptake calculated using neck background (3.96±2.76% vs. 3.74±2.73%, P < 0.001). Thyroid uptake values obtained using the two background measurement methods were correlated (r = 0.994, P < 0.001). Bland–Altman analysis demonstrated a mean bias of 0.223%, with 95% limits of agreement ranging from −0.358% to 0.804%. Female patients showed a tendency toward a larger difference in thyroid uptake values than male patients, although the difference was not statistically significant (P = 0.121). Conclusion: Although neck background measurement using a B-type filter resulted in significantly higher background counts and slightly different thyroid uptake values compared with mid-thigh measurement, the two methods demonstrated overall agreement. Nevertheless, mid-thigh background measurement may provide a more stable assessment by reducing the influence of adjacent thyroid-related structures. These findings suggest that mid-thigh background measurement may be useful for more consistent thyroid uptake evaluation in 99mTc-pertechnetate studies.
Keywords

99mTc-pertechnetate, Thyroid uptake, Background activity, B-type filter, Nuclear medicine