Analysis of Image Quality on CT and PET following the Application of the iMAR Algorithm in Patients with Metallic Implants

Department of Nuclear Medicine, Asan Medical Center, Seoul, Korea

Correspondence to Jae-Kwang Ryu, Department of Nuclear Medicine, Asan Medical Center, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, Republic of Korea. Tel: +82-2-3010-5421, E-mail: huhjoon11@naver.com

Volume 30, Number 2, Article 24, November 2026. Korean J Nucl Med Technol 2026;30(2):24. https://doi.org/10.12972/kjnmt.2026.30.2.24
Received on November 02, 2026, Revised on November 13, 2026, Accepted on November 16, 2026, Published on November 30, 2026.
This is an Open Access article distributed under the terms of the Creative Commons CC BY 4.0 license (https://creativecommons.org/licenses/by/4.0/) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: Metallic implants can lead to metal artifacts in PET/CT imaging, which may affect CT-based attenuation correction and PET quantification. The aim of this study was to evaluate differences in quantitative parameters of CT and PET images following the application of implant-specific iMAR(Iterative Metal Artifact Reduction) algorithms. Materials and Methods: This retrospective study included 81 patients who underwent 18F-FDG PET/CT scans; it analyzed a total of 90 cases of metallic implants. Representative CT and PET images before and after iMAR application were compared visually. For quantitative analysis, CT HUmean, PET SUVmean, and SUVmax were measured, while liver SUVmean was assessed separately. Results: A visual comparison of the representative images showed that the metal artifacts decreased in CT images after iMAR application, but no obvious visual differences were observed in PET images. In the quantitative analysis, for all three implant types, CT HUmean, PET SUVmean, and SUVmax in the bright band regions decreased significantly after iMAR application, while they increased significantly in the dark band regions. In contrast, no statistically significant differences were observed in liver SUVmean. Conclusion: Implant-specific iMAR application resulted in significant differences in CT HUmean, PET SUVmean, and SUVmax in metal artifact regions. These results suggest that the reconstruction conditions of CT images used for PET attenuation correction may affect the PET quantification values around the metal artifact area. Therefore, when interpreting the quantitative values of areas adjacent to metal in PET/CT images, the application of iMAR must be considered.
Keywords

PET/CT, iMAR, Metal Artifact, HU(Hounsfield Unit), SUV(Standardized Uptake Value)